INSTRUCTION MANUAL
Guidance and manufacturer's declaration- electromagnetic immunity-
For EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The model MD300K2 Pulse Oximeter is intended for use in the electromagnetic specified
below. The customer of the user of the MD300K2 Pulse Oximeter should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Conducted
RF IEC
3Vrms 150kHz to
61000-4-6
80kHz
Radiated RF
3V/m 80MHz to
IEC 61000-4-
2.5GHz
3
NOTE1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
3V
3V/m
- 31 -
Electromagnetic environment-
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the model MD300K2 Pulse
Oximeter, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
150KHz to 80 MHz
80MHz to 800MHz
800MHz to 2.5GHz
Where P is the maximum output
power rating of the transmitter
in Watts (W) according to the
transmitter manufacture and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,a should
be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
MD300K2