MD300K2
Guidance and manufacturer's declaration - Electromagnetic Immunity-
Guidance and manufacturer's declaration- electromagnetic immunity
The model MD300K2 Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The customer or the user of the model MD300K2 Pulse Oximeter should
assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-4
Electrostatic
transient/burst
IEC 61000-4-4
Surge IEC 61000-
4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-1
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
For all Equipment and Systems
IEC 60601 test level
±6kV contact ±8kV air
±2kV for power supply
lines ±1kV for input/
output lines
±1kV differential mode
±2kV common mode
<5% UT ( >95%dip
in UT) for 0.5 cycles
40% UT (60%dip in
UT) for 5 cycles 70%
UT( 30%dip in UT) For
2 5 cycles <5% UT (
>95%dip in UT) for 5
sec
3A/m
Compliance level
±6kV contact ±8kV
air
±2kV for power
supply lines ±1kV
for input/output
lines
±1kV differential
mode ±2kV
common mode
<5% UT ( >95%dip
in UT) for 0.5 cycles
40% UT (60%dip
in UT) for 5 cycles
70% UT (30%dip in
UT) For2 5 cycles
<5% UT ( >95%dip
in UT) for 5 sec
3A/m
- 30 -
PULSE OXIMETER
Electromagnetic
environment-guidane
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the model MD300K2
Pulse Oximeter requires
continued operation during
power main Interruptions,
it is recommended that the
model MD300K2 Pulse
Oximeter be powered
from an uninterruptible
power supply or a battery.
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.