MD300K2
APPENDIX A SPECIFICATIONS
Notes:
Specifications may be changed without prior notice.
The circuit diagrams, the list of components, the illustration of diagrams, and the detailed
rules of calibration, are provided exclusively to professional personnel authorized by our
company.
The equipment has been calibrated, users do not need to calibrate. In order to ensure
the accuracy of the probe, please change the probe once a year. Make sure that the type of
probe need to be specified.
Display
Data: SpO
, PR, pulse column, PI, SpO
2
Others: Information such as connection status of probe and power low indication.
Alarm
Alarm: SpO
and pulse rate value, probe off, battery exhausted
2
Alarm mode: audio alarm, visual alarm and information
Alarm limits range: SpO
Default limits: SpO
SpO
2
Measurement range: 0~100%
Resolution:
1%
Accuracy:70%~100%; ±2%; <70% unspecified
Pulse Rate
Measurement range: 30bpm~250bpm
Resolution: 1 bpm
Accuracy: ±2 bpm or 2% (The larger are applied)
Perfusion Index
Display range: 0.1%~20%
Probe LED Specifications
RED
IR
70%~100%, PR 30bpm~250bpm
2
High 100%, low 90%; PR High 100bpm; low 60bpm
2
plethysmogram
2
Wavelength
660±3nm
905±10nm
- 24 -
PULSE OXIMETER
Radiant Power
3.2mW
2.4mW