SETTING

5.1 System Setting

Pick and enter the [System Setting] interface from the main menu. And then press the
Navigation buttons to select different item to set.
Measuring
SpO
100
090
Alarm Volume: you can adjust the value of alarm volume, there are 7 levels, and the default
level is 3.
Alarm Pause: there are two modes, 60s and 120s, and the default mode is 120s. The device
audibly alarm upon new alarm conditions.
Beep Tone: the level is from 0 to 7, and the default level is 3.
Backlight Setting: the level of brightness is from 1 to 7, and the default level is 3.
Language: English, French, German, Spanish, Italian, Japanese, Russian and Chinese.
Screen Sleep Mode: 1minute, 10 minutes, 30 minutes, screen always on, and the default is
1 minutes
Charging management: Charging activated, Charging stop.
ID setup: ① press OK button, ② press Navigation button to change the number, ③ press OK
button to confirm.
Factory Default: recover to factory reset.
Note:
1. Every time enter into the Alarm Volume and Alarm Pause, you should input the password,
refer to Alarm Setting.
2. The ID range is 1~127.
120s
%
PR
bpm
2
98
70
100
060
System Setting
Alarm Volume
Alarm Pause
Beep Tone
Backlight Setting
Language
Screen Sleep Mode
OK
12:00
Fig.5.1
Measuring
SpO
%
2
98
100
090
System Setting
Charging Management
5
ID setup
120s
Factory Default
5
5
Back
OK
- 12 -
PULSE OXIMETER
120s
PR
bpm
70
100
060
01
12:00
Back
Fig.5.2

Publicité

Chapitres

loading

Produits Connexes pour Medipost-Nutristoma MD300K2