INSTRUCTION MANUAL
Warnings!
Use only SpO
probes provided by manufacturer. Other SpO
2
performance.
Do not use an SpO
Excessive patient movement may cause inaccurate measurements.
Tissue damage can be caused by incorrect operation or misusing probe; for example,
by wrapping the probe too tight. Inspect the probe site to ensure the skin's integrity and
the adhesion position of the probe is correct. More frequent inspection should be taken if
necessary.
Loss of pulse signal can occur in any of the following situations:
a) The probe is too tight;
b) There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp,
or sunlight; Pulse sensor should obviate the light source, e.g. radial lamp or infrared lamp.
c) A blood pressure cuff is inflated on the same extremity as the one to which an SpO
is attached.
After measurement, please take off your finger and press and hold the power button to turn
off the device.
probe with exposed optical components.
2
- 11 -
MD300K2
probes may cause improper
2
sensor
2