MD300K2
8 MAINTENANCE AND REPAIR
Warnings!
The advanced circuit inside the oximeter does not require periodic calibration and
maintenance, except replacing the batteries.
Don't open the cover of oximeter or repair electronic circuits. Its open will cause the
damage of the device and the annulment of the guarantee.
8.1 Maintenance
Use only the substances approved by us and methods listed in this chapter to clean
or disinfect your equipment. Warranty does not cover damage caused by unapproved
substances or methods.
We make no claims regarding the efficacy of the listed chemicals or methods as a means
for controlling infection. For the method to control infection, consult your hospital's Infection
Control Officer or Epidemiologist. Keep your equipment and accessories free of dust and dirt.
To avoid damage to the equipment, follow these rules:
Always dilute according the manufacturer's instructions or use lowest possible
concentration.
Do not immerse part of the equipment into liquid.
Do not pour liquid onto the equipment or accessories.
Do not allow liquid to enter the case.
Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such
as acetone or acetone-based cleaners).
Caution: If you spill liquid on the equipment or accessories, contact us or your service
personnel.
Note: To clean or disinfect reusable accessories, refer to the instructions delivered with the
accessories.
8.2 Safety Checks
Before every use, or after your pulse oximeter has been used for 6 to 12 months, or whenever
your pulse oximeter is repaired or upgraded, a thorough inspection should be performed by
qualified service personnel to ensure the reliability. Follow these guidelines when inspecting
the equipment:
Make sure that the environment and power supply meet the requirements.
Inspect the equipment and its accessories for mechanical damage.
Make sure that only specified accessories are applied.
Inspect if the alarm system functions correctly.
Make sure that the batteries meet the performance requirements.
- 20 -
PULSE OXIMETER