ture. For this, use the SURFACE TEMP mode.
Measurement range for SURFACE TEMP mode: 0°C – 60°C (32°F - 140°F).
By selecting the ROOM Mode, it will display the ambient temperature of
the room you are in.
Press the MODE button for one second, the screen will display: BODY
for the body measurement temperature. Press the MODE button again
to switch to SURFACE TEMP or ROOM mode.
Press the MODE button again to switch to Surface Temp or Room mode.
Note: The ThermoFlash® LX-26 EVOLUTION is automatically set to BODY.
Important: The surface temperature differs from the internal body
temperature. To obtain the internal temperature, always use the BODY
mode. Please make sure to select the BODY mode for an internal tem-
perature reading and the SURFACE TEMP mode or the ROOM mode for
an external area reading (bottle, bath, room...).
2. Choosing the temperature unit
Press the °C/°F button to select the temperature unit. The symbol
of the chosen temperature unit is displayed (°C for Celsius de-
grees and °F for Fahrenheit degrees).
3. Data memory
To display the last temperature measurement, press the MEM button.
You will obtain the last temperature measurement. To switch on
the last measurement, press the °C/°F button.
The number indicated in interval of two measures corresponds
to the number of measurements. Press on the button MODE for
review of the previous measurements.
To exit data memory, press the measure key. The Thermoflash®
LX-26 EVOLUTION switches off after 5 seconds.
4. Alarm setup
The alarm threshold default value is 38°C (100.4°F). It cannot
be changed. In case of higher temperatures than 38°C, a signal
beeps while measuring the temperature.
5. Changing the batteries
Display: when the LCD screen displays
Operation: Open the lid and change the batteries, taking great
care with the correct positioning. A mistake with this could cause
damage to the apparatus and compromise the guarantee of your
ThermoFlash® LX-26 EVOLUTION. Never use rechargeable bat-
teries. Use only batteries for single usage.
VIII. TECHNICAL CHARACTERISTICS & PRECISION
1. Normal conditions of use operating temperature: 10°C ~ 40°C
(50°~ 104°F). Normal conditions of storage and transportation
temperature: -20°C ~ 50°C (-68°~ 122°F). Humidity rate: ≤ 85%.
22
Visiomed - LX-26E - UK - 092014
, the battery is used.
2. Power: DC3 V (2 batteries LR6).
3. Size: 155 x 90 x 38 mm.
4. Weight: 220g.
5. Display Resolution: 0.1°C (0.1°F).
6. Measuring range:
In Body mode: 32°C ~ 42.9°C (90°F – 109°F).
In Surface Temp mode: 0°C ~ 60°C (32°F to 140°F).
7. Precision: From 36°C to 39°C (96.8°F to 102.2°F) = +/-0.2°C/°F.
8. Consumption: ≤ 50mW.
9. Measuring distance: 5 cm – 8 cm (2 in – 3.14 in).
10. Automatic stop: 5 sec.
Longevity use
The ThermoFlash® is guaranteed for 40.000 readings.
IX. ADVICE
- The protective glass over the lens is the most important and
fragile part of the thermometer, please take great care of it.
- Clean the glass with cotton fabric, wet with water or 70° alcohol.
- Do not use other batteries than mentioned batteries, do not re-
charge non rechargeable batteries, do not throw in fire.
- Remove the batteries when thermometer is not used for an
extended period of time.
- Do not expose the thermometer to sunlight or water.
- An impact will damage the product.
X. ACCESSORIES SUPPLIED
User Manual, Fast learning user manual, Guarantee card, Batteries
supplied (LR6).
XI. TROUBLESHOOTING
If you have one of the following problems while using your Ther-
moFlash® LX-26 EVOLUTION, please refer to this breakdown ser-
vice guide to help resolve the problem. If the problem persists,
please contact our customer service at +33 892 350 334.
- The screen displays temperature superior to 95°: the tempe-
rature is in Fahrenheit. Change the measurement unit to Celsius.
- The screen displays the body temperature inferior to 32°C
(89.6°F): to measure a body temperature, the BODY mode must
be selected. If you are on Surface Temp mode, the 32°C (89.6°F)
temperature displayed is showing the external temperature
that your body releases.
- The screen displays the message HI: when using the Ther-
moFlash® LX-26 EVOLUTION, the message HI can show on the
Visiomed - LX-26E - UK - 092014