Sterilisation Parameter Settings; Configuration Parameter Settings - VWR VAPOUR-Line eco 25 Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
11. Press START to start the cycle
You can stop the sterilisation cycle at any time by pressing the STOP button.
WARNING! STOP button is an emergency button. Pushing it will abort the cycle and will open the
valves. We don't recommend you push it unless a problem occurs during the cycle.
The same thing will happen if the instrument is turned off by the main switch.

Sterilisation parameter settings

1. Turn on the autoclave by the main switch (Pos. 1 page 7).
2. After the initial presentation the display will show the Program 1 (P1).
P
1
3. To modify this programme press the ENTER button, or press the  or  buttons until the programme to be
modified is displayed and then press ENTER.
4. An arrow will be displayed next to the sterilisation temperature, press the  or  buttons until the desired
sterilisation temperature is displayed and then press ENTER.
P
1
5. An arrow will be displayed next to the sterilisation time, press the  or  buttons until the desired
sterilisation temperature is displayed and then press ENTER.
6. The modified data is automatically stored in the memory.
7. Press the START button to start the sterilisation cycle.
NOTE:
Program 5 has a double cycle making it possible to set two independent temperatures at the same time.

Configuration parameter settings

Press the MENU button to modify the instrument configuration parameter settings.
There are seven sub menus:
• "Printer" - modifies the printer settings
• "Delay" - sets a delayed start of the sterilisation cycle
• "Print last cycle" - prints the last sterilisation cycle completed
• "Password" - sets a password to prevent unauthorised modification of the sterilisation and instrument
settings
• "Date and time" - modifies the autoclave date and time
• "LCD contrast" - modifies the adjustment of the display contrast