Check that the instrument passes power-on self tests

Power-on self tests verify that all instrument modules are working correctly after power up.
Prerequisite: For the Low Profile instruments, connect a monitor to a video output on the rear panel and connect a mouse to a
USB port.
1.
Power on the instrument and wait until the instrument screen appears on the monitor.
2.
Select Utility > Self Test from the top-edge Menu bar to open the Self Test configuration menu.
3.
Check that the status of all power-on self tests are Passed.
If one or more power-on self tests shows Failed:
a.
Power cycle the instrument.
b.
Tap Utility > Self Test. If one or more power-on self tests still shows Failed, contact Tektronix Customer Support.

Connecting signals to the instrument

Cables connect the instrument to your device under test (DUT). Use a cable that best matches your signal measurement needs.
Figure 2: Connecting SMA cables to the LPD64
Connecting SMA cables (LPD64)
Connect SMA cables to the instrument. Hand tighten, then torque the cable connectors to 56 N-cm (5 in-lb).
LPD64 Installation and Safety Manual
Installing your instrument
5