Table 2
Guidance and manufacturer's declaration – electromagnetic Immunity
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
ENGLISH
IEC 60601-1-2
Test level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
0% U
; 0,5 cycle. At 0°, 45°, 90°,
T
135°, 180°, 225°, 270° and 315°.
0% U
; 1 cycle and 70% U
T
25/30 cycles; Single phase: at 0°.
0% U
; 250/300 cycle
T
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80% AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
Compliance level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
0% U
; 0,5 cycle. At 0°, 45°, 90°,
T
135°, 180°, 225°, 270° and 315°.
;
0% U
; 1 cycle and 70% U
T
T
25/30 cycles; Single phase: at 0°.
0% U
; 250/300 cycle
T
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80% AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
56
;
T