9-2 Detaching and reattaching the door
9-2-1 Detaching the door
Slide the detachment lever (lower right corner) inward then pull the door out to remove
(Fig. 22).
9-2-2 Reattaching the door
1) Fit the installation hole on the door into the pin on the front left side (Fig. 23).
2) Align the other installation hole on the door right side with the pin while pulling the
detachment lever. Release the detachment lever. A 'click' sound will indicate that the
door is secure (Fig. 24).
• Attach or detach the door on a level surface. During attachment or detachment, be careful not to drop the door or the Unit.
• Do not soak the door in alcohol and do not leave an alcohol-immersed cloth in contact with the door for a long period. Bleaching
CAUTION
or door deformation may occur.
9-3 Replacing Mist Filter
Open the door and remove the used Mist Filters. Replace with new Mist Filters as shown in
Fig. 25.
• Replace immediately with new Mist Filters when signs of debris appear.
• Dispose of used Mist Filters as medical waste according to the laws and
CAUTION
regulations in your country.
9-4 Replacing the Oil Absorber Sheet
Open the door and remove used Oil Absorber Sheet. Place new Oil Absorber Sheet at the
bottom of the lubricating chamber of the Unit as shown in Fig. 26.
• Replace immediately with a new Oil Absorber Sheet when signs of debris
appear.
CAUTION
• Dispose of used Oil Absorber Sheets as medical waste according to the laws
and regulations in your country.
Pin
Pin
Installation hole
Installation hole
12
Detachment lever
Detachment lever
Installation hole
Installation hole
Fig. 23
Detachment lever
Detachment lever
Fig. 22
Pin
Pin
Fig. 24
Mist Filter
Mist Filter
Fig. 25
Oil Absorber Sheet
Oil Absorber Sheet
Fig. 26