Transport
CAUTION
Damage to the sensor unit
The sensor unit may be damaged by the ingress of liquid.
► Make sure that no liquid penetrates into the gas inlet openings and the gas
outlet opening on the sensor unit.
► Clean carefully as required. Do not use detergents or solvents.
CAUTION
Damage to the activated carbon filter
Detergents and solvents can stress and potentially overstress the internal
activated carbon filter.
► Do not use detergents or solvents near an operating sensor unit.
● Dry the sensor unit with a cloth.
6
Transport
The sensor unit contains lithium batteries. During transport of the sensor unit,
and particularly during air transport, observe the relevant safety regulations for
lithium batteries.
Do not transport the sensor unit at below -4 °F (-20 °C) or above 122 °F
(+50 °C) as it could become damaged.
NOTICE
Observe the safety instructions for the control unit!
The safety instructions for the control unit contain separate information on
storage and must be observed.
7
Storage
–
Store in a dry environment free of pollutants.
–
Dräger recommends charging the sensor unit and the control unit during
storage.
–
Dräger recommends checking the charge status of the energy supply at
least every 3 weeks if the sensor unit or the control unit is not left on charge.
–
Do not store the sensor unit at below -4 °F (-20 °C) or above 122 °F
(+50 °C) as it could become damaged.
NOTICE
Observe the safety instructions for the control unit!
The safety instructions for the control unit contain separate information on
storage and must be observed.
WARNING
Risk of explosion!
Discharging isolated capacitances can create ignitable sparks.
► When storing the device, make sure that it is grounded so that no isolated
capacitance can be electrostatically charged.
To keep power consumption low in the heat-up phase, and to keep the heat-
up phase as short as possible, store the sensor unit at room temperature.
8
Disposal
This product must not be disposed of as household waste. This is indi-
cated by the adjacent icon.
You can return this product to Dräger free of charge. For information
please contact the national marketing organisations and Dräger.
®
Dräger X-pid
8x00 Sensor Unit