Hach AN-ISE sc Mode D'emploi page 16

Masquer les pouces Voir aussi pour AN-ISE sc:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 59
Section 5
5.3 Replace the sensor cartridge
16
Maintenance
The sensor cartridge is replaced as described below and in
Figure 15 on page
1. Replace the cartridge using menu item
2. Clean the probe and thoroughly dry the sensor cartridge and
3. Loosen the 4 socket head screws.
The sensor cartridge must point downward such that no water can run into
the probe adapter. Pay attention to the contacts between the probe and
the sensor cartridge. These contacts must remain dry.
4. Pull the sensor cartridge out of the probe adapter and discard
5. Make sure that a new black gasket is installed every time the
6. Insert the new sensor cartridge into the probe adapter. Observe
7. Secure the sensor cartridge with the 4 socket head screws.
8. The sensor code (calibration data) is automatically read for
17.
or
or
AN-ISE SC
AISE SC
CARTRIDGE REPLACEMENT
probe adapter.
N O T I C E
the old sensor cartridge as per the applicable regulations.
sensor cartridge is replaced. Before the gasket is installed,
clean the surface that faces the cartridge and the groove for the
gasket.
the marker hole on the sensor cartridge flange and the probe
adapter.
instruments with automatic recognition (LXG440.99.x000x). For
instruments with non-automatic recognition
(LXG440.99.x001x), enter the new sensor code manually (refer
to certificate).
NISE SC > MAINTENANCE > SERVICE >
.
.
English

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Aise scNise sc

Table des Matières