PRODUCT SPECIFICATIONS
Urgent PC Stimulator
1. Design Features:
» Lightweight, ergonomic, handheld design
» Electronic touch pad controls
» Raised and embossed buttons to provide tactile feedback
» LCD status screen to provide operational status
» One-way fit connection site for Lead Set
2. Electrical Current Settings:
Current
Current
Setting
(in mA)
0
0
1
0.15
2
0.5
3
1.0
4
1.5
5
2.0
6
2.5
7
3.0
8
3.5
9
4.0
» Pulse characteristics:
◦ Fixed pulse frequency of 20 Hz
◦ Pulse width of 200 µseconds
◦ Square waveform
» Resistance of 500-4000 Ohms.
8
Current
Current
Setting
(in mA)
10
4.5
11
5.0
12
5.5
13
6.0
14
6.5
15
7.0
16
7.5
17
8.0
18
8.5
19
9.0
Button Controls
The Stimulator is controlled by raised buttons on the device. These
buttons are:
Power button – turns power on and off
Note: To turn the power on or off, the user is required to depress
the Power button for approximately 2 seconds. This is designed
to protect the Stimulator from inadvertent status changes.
Test (yellow) button – begins Test mode
Upon entering Test mode, the default current setting will be 0
(0mA). At the completion of the Test mode, the final current
setting shall be the baseline setting in Therapy mode.
Note: To activate Test mode, the user is required to depress the
Test mode button for approximately 2 seconds. This is designed
to protect the Stimulator from inadvertent status changes.
Therapy (green) button – begins Therapy mode
The default current setting for Therapy mode will be the final
current setting in Test mode. However, the Current Adjustment
button may be used to increase or decrease the current level at
any time.
Note: To activate Therapy mode, the user is required to depress
the Therapy mode button for approximately 2 seconds. This is
designed to protect the Stimulator from inadvertent status changes.
Stop (red) button – stops flow of current in Test or Therapy
mode
Note: If treatment is stopped or interrupted during Therapy mode,
the remaining Therapy mode time is displayed. Once stopped,
the treatment session will need to be restarted, beginning with
Test mode. Depending on when the session was interrupted, a
new Lead Set may be required.
Current Adjustment button – increases or decreases current
Note: The current is adjustable in both Test and Therapy modes.
Note: if the Stimulator is not in Test or Therapy mode and no button is
pressed for 15 minutes, the Stimulator will power down.
9