TMR31, TMR35
Long-term stability of electronics
≤ 0.1 K (0.18 °F)/year or ≤ 0.05%/year
Data under reference conditions. % relates to the set span. The larger value applies.
Influence of ambient temperature (temperature drift)
• Pt100 resistance thermometer:
T
= ±(15 ppm/K * (full scale value of measuring range + 200) + 50 ppm/K * set measuring range) * Δ ϑ
d
Δ ϑ = deviation of ambient temperature from the reference operating condition.
Influence of load
± 0.02%/100 Ω. Specifications refer to the full scale value of the measuring range.
Transmitter response time
1 s
Sensor response time (measured as per IEC 60751, in water flowing at 0.4 m/s (1.3 ft/s)
t
< 1.0 s
50
Influence of supply voltage
≤ ±0.01%/V deviation from 24 V. Specifications in percent refer to the full scale value of the measuring range.
Self-heating
Negligible small
Sensor current
< 0.6 mA
Endress+Hauser
t
< 2.0 s
90
Technical data
33