Wiring - Endress+Hauser Easytemp TMR31 Instructions D'opération

Masquer les pouces Voir aussi pour Easytemp TMR31:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TMR31, TMR35
4

Wiring

TMR35: Electrical connection cables must comply with 3-A sanitary standard, must be
smooth, corrosion resistant and cleanable.
NOTICE
To prevent damage to the device
Do not tighten the M12 plug too much, in order to prevent damage to the device.
M12x1
Fig. 3: Pin assignment of the 4-wire M12 plug for output signal and power supply.
A: with transmitter, M12 plug, 4-pin
Pin 1: power supply 10 to 35 V DC; 4 to 20 mA (cable connection, wire color brown = BN)
Pin 2: connection of PC configuration cable - shortened pin (cable connection, wire color white = WH)
Pin 3: power supply 0 V DC; 4 to 20 mA (cable connection, wire color blue = BU)
Pin 4: connection of PC configuration cable - shortened pin (cable connection, wire color black = BK)
B: without transmitter, Pt100, 4-wire connection
Endress+Hauser
1 (BN) +
2 (WH) nc
3 (BU) -
4 (BK) nc
A
4
1
10...35 VDC
(4-20 mA)
3
0 V
(4-20 mA)
2
B
Wiring
J
Pt100
A0020176
27

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Easytemp tmr35

Table des Matières