SETTING UP THE SCALES

Attach the power supply module to the connector on the side of the indicator.
Press the [On/Off] key. The software revision number will be displayed
followed by a self-test showing all digits before the zero is displayed along with
the unit of weight that was selected last.

3.4 CONNECTIONS

3.4.1 CONNECTION OF LOAD CELL

Please see figure 1
2
5
7
Note:
For 4 wire load cell, connect the load cell +Excitation and +Sense together at the
connector and -Excitation and -Sense together at the connector.
1
4
3
6
As viewed from the back of the indicator
6: Pin +IN
+Signal
7: Pin –IN
-Signal
3: Pin AGND
Shield
1: Pin +E,
+Excitation
2: Pin +S
+Sense
4: Pin –E
-Excitation
5: Pin -S
-Sense