12.3.3
F3 – CAPACITY/ WEIGHING UNIT/INCREMENT
To enter this parameter, press the [Z/T] key when "F3 CAP" is shown.
The display will show the current capacity and default weighing unit.
Enter the numeric values using the [Unit/] and [Pcs/] keys.
The scale will
check the number of division's n = maximum/increment is less than 30,000 divisions.
(3000 divisions for EC approved versions).
To select the weighing unit press the [Unit/]until the default weighing unit is
flashing, either kilogram (kg) or pounds (lb).
Press [Z/T] to accept the displayed value.
On non-approved versions the display then lets you select the increment, "InC 2"
For example 100kg x 0.01kg the increment is 10grams, but the last digit increments
by 1.
The display will show the current increment value as used with the current decimal
point position.
Press the [Pcs/] key to change the value. Select from 1, 2, 5, 10, 20 or 50
Not all increments may be available for the capacity you selected.
For EC Approved versions the scale will determine the increment that maintains the
number of divisions to be 3000 or less.
Press [Z/T] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.
© Adam Equipment Company 2010