G
®
AERASGARD
KCO
Patented quality product (patent no. DE 10 2014 010 719.1)
The maintenance-free, microprocessor-controlled AERASGARD
used to detect CO
content as well as gas temperature. The measurement signals are converted to standard signals of 0 -10 V.
2
The CO
content of the air is measured using an optical NDIR sensor (non-dispersive infra-red technology). The detection range of the CO
2
calibrated for standard applications such as monitoring residential rooms and conference rooms. Room ventilation on an as-needed basis, improved
well-being and customer benefit, increased comfort as well as reduced operating costs through energy conservation are just some of the benefits of
®
employing AERASGARD
CO
2
A measuring system based on NDIR (non-dispersive infra-red technology) for CO
range of wavelengths of light radiated by the source is damped and absorbed by CO
the receptor.
TECHNICAL DATA
Voltage supply:
Average power consumption:
CARBON DIOXIDE (CO
)
2
Sensor, CO
:
2
Measuring range, CO
:
2
Output, CO
:
2
Measuring accuracy, CO
:
2
Temperature dependence:
Pressure dependence:
Long-term stability:
Gas exchange:
TEMPERATURE
Measuring range, temperature:
Output, temperature:
Warm up time:
Ambient temperature:
Humidity:
Response time:
Electrical connection:
Enclosure:
Dimensions:
Cable gland:
Protective tube:
Process connection:
Protection class:
Protection type:
Standards:
- SD - TYR2 ⁄ KTM - CO
2
sensors.
24 V AC ⁄ DC
< 3 VA at 24 V DC
optical NDIR sensor
(non-dispersive infra-red technology)
dual-beam measuring system
with automatic calibration
multi-range switching
(selectable via DIP switches)
0...2000 ppm ⁄ 0...5000 ppm
0 -10 V
up to 2000 ppm: ± 70 ppm plus 5 % of measured value
up to 5000 ppm: ± 70 ppm plus 10 % of measured value
approx. 10 ppm ⁄ K at +25 °C
± 1.5 % of measured value ⁄ kPa (referred to standard pressure)
± 2 % of final value ⁄ year
by diffusion,
minimum flow rate 0.2 - 0.5 m/s
0...+50 °C
0 -10 V
approx. 1 hour
+10...+50 °C
< 95 % r. H., non-precipitating air
approx. 2 minutes
0.14 - 1.5 mm
2
,
via screw terminals
plastic, material polyamide, 30 % glass-globe reinforced,
with quick-locking screws (slotted ⁄ Phillips head combination),
colour traffic white (similar to RAL 9016)
126 x 90 x 50 mm (Tyr 2)
M 16 x 1.5; including strain relief, exchangeable
PLEUROFORM
TM
, material polyamide (PA6), with torsion protection,
Ø 20 mm, NL = 202.5 mm, v
via flange made of plastic
(included in scope of delivery)
III (according to EN 60 730)
IP 65 (according to EN 60 529) enclosure only!
CE conformity,
electromagnetic compatibility according to EN 61 326,
EMC Directive 2014 ⁄ 30 ⁄ EU
- SD - TYR2
2
®
KCO
- SD - TYR2 or KTM - CO
2
measurement consists of a light source and a receptor. A certain
2
molecules in the measured section. This damping is detected by
2
= 30 m/s (air)
max
- SD - TYR2 is designed for duct installation and is
2
Rev. 2017 - V35 GB
sensor is
2