SECTION 11 – TROUBLESHOOTING
11.1

CALIBRATION ERROR CODES

If an error code is displayed during calibration it will be accompanied by a Clipboard icon and a
number to indicate the cause of the error. The first time that an error is displayed, it will be possible
to repeat the erroneous measurement. If the same error message is shown after the repeat
measurement, the complete calibration process will need to be repeated. The table below shows
the error codes:

11.1.1 Calibration Procedure Error Codes

Error Code
Symbol
1
2
3
4
5
6
7
11.2

TROUBLESHOOTING GUIDE

Issue
Unable to achieve
zero absorbance or
100% transmittance
when calibrating
Unable to achieve
a reading when
measuring a sample
11.3

TECHNICAL SUPPORT

Jenway have a dedicated Technical Support team made up of experienced scientists who are on
hand to help with any applications advice and questions you may have about our products or how
to use them. If you require any technical or application assistance please contact the team at:
E-mail:
cptechsupport@coleparmer.com
Phone: +44 (0)1785 810433
28
Issue
Absorbance range of calibration blanking solution for calibration
steps is too large
Background absorbance (330nm) of calibration solution is too
large (first calibration solution)
Background absorbance (330nm) of calibration solution is too
large (second calibration solution)
Variance in pair of calibration solution results is too large
Calculated calibration pathlength factors are too large
Absorbance range of calibration blanking solution for verificati on
steps is too large
Calibrated verification absorbances are outside of acceptable
limits
Solution
Ensure that there is not a sample on the micro volume accessories read head.
Ensure the instrument lid is closed before and during the calibration.
Ensure the lamp is working – if the lamp has failed please request service
assistance.
Ensure the correct path length is being used.
Ensure the sample isn't so dense so that light is not transmitted through the
sample.
Ensure the lamp is working.