RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND ICARE IC100
Icare ic100 (TA011) is intended for use in an electromagnetic environment in which radiated RF-
disturbances are controlled. The customer or the user of the Icare ic100 (TA011) can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and Icare ic100 (TA011) as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
Separation distance according to frequency of transmitter, m
output power of
150 kHz to 80 MHz
transmitter
d= 1.2 √P
(W)
0.01
NOT APPLICABLE
0.1
NOT APPLICABLE
1
NOT APPLICABLE
10
NOT APPLICABLE
100
NOT APPLICABLE
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d= 1.2 √P
d= 2.3 √P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.2
12
23
English
13