GUIDANCE AND MANUFACTURER'S DECLARATION–ELECTROMAGNETIC EMISSIONS
Icare ic100 (TA011) is intended for use in the electromagnetic environment specified below.
The user of the Icare ic100 (TA011) should assure that it is used in such an environment.
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
flickering emissions
IEC 61000-3-3
GUIDANCE AND MANUFACTURER'S DECLARATION–ELECTROMAGNETIC IMMUNITY
Icare ic100 (TA011) is intended for use in the electromagnetic enviroment specified below.
The customers or users of Icare ic100 (TA011)) should assurre that it is used in such an enviroment
Immunity test
IEC 60601 Test level Compliance level
Electrostatic discharge
± 6 kV contact
(ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
Transients/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV for line(s) to
IEC 61000-4-5
line(s)
±2 kV for line(s) to
earth
Voltage dips, short
<5 % UT (>95 %
interruption and
dip in UT) for
voltage variations on
0,5 cycle
power supply lines
40 % UT (60 % dip
IEC 61000-4-11
in UT) for 5 cycles
70 % UT (30 %
dip in UT) for
25 cycles
<5 % UT (>95 %
dip in UT) for 5 s
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
Group 1
Icare ic100 (TA011) is battery operated and
use RF energy only for its internal function.
Therefore, its RF emissions are low and are
not likely to cause any interference in nearby
equipment.
Class B
Icare ic100 (TA011) is suitable for use in
all establishments, including domestic
NOT APPLICABLE
establishments and those directly connected to
public low-voltage power supply network that
NOT APPLICABLE
supplies buildings used for domestic purposes
± 6 kV contact
± 8 kV air
NOT
APPLICABLE
NOT
APPLICABLE
NOT
APPLICABLE
3 A/m
Electromagnetic environment-
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %
NOT APPLICABLE
NOT APPLICABLE
NOT APPLICABLE
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
English
11