*ESE?
*ESR?
*IST?
*OPC
*OPC?
*PRE <nrf>
*PRE?
QER?
*RST
*SRE <nrf>
*SRE?
*STB?
*WAI
Miscellaneous Commands
*IDN?
ADDRESS?
*TST?
*TRG
Error Messages
Each error message has a number; only this number is reported via the remote control interfaces.
Error message numbers are not displayed but are placed in the Execution Error Register where
they can be read via the remote interfaces, see Status Reporting section.
34
Returns the value in the Standard Event Status Enable Register in <nr1>
numeric format. The syntax of the response is <nr1><rmt>
Returns the value in the Standard Event Status Register in <nr1> numeric
format. The register is then cleared. The response is <nr1><rmt>. See
Status Reporting section for details.
Returns ist local message as defined by IEEE Std. 488.2. The syntax of the
response is 0<rmt>, if the local message is false, or 1<rmt>, if the local
message is true.
Sets the Operation Complete bit (bit 0) in the Standard Event Status
Register. This will happen immediately the command is executed because of
the sequential nature of all operations.
Query Operation Complete status. The response is always 1<rmt> and will
be available immediately the command is executed because all commands
are sequential.
Set the Parallel Poll Enable Register to the value <nrf>.
Returns the value in the Parallel Poll Enable Register in <nr1> numeric
format.
The syntax of the response is <nr1><rmt>
Query and clear Query Error Register. The response format is nr1<rmt>
Resets the instrument to the remote control default settings with the
exception of all remote interface settings and Output state at power-on
setting. (see Remote Operation Defaults paragraph in the Remote Interface
Operation section)
Set the Service Request Enable Register to <nrf>.
Returns the value of the Service Request Enable Register in <nr1> numeric
format. The syntax of the response is<nr1><rmt>
Returns the value of the Status Byte Register in <nr1> numeric format. The
syntax of the response is<nr1><rmt>
Wait for Operation Complete true. As all commands are completely executed
before the next is started this command takes no additional action.
Returns the instrument identification. The exact response is determined by
the instrument configuration and is of the form <NAME>,<model>,<Serial
No.>, <version><rmt> where <NAME> is the manufacturer's name, <model>
defines the type of instrument, Serial No. is the unique serial number and
<version> is the revision level of the software installed. The format of
<version> is
X.xx – Y.yy where X.xx is the revision of the main instrument firmware and
Y.yy is the revision of the Interface board firmware. The Interface firmware is
field updateable by the user via the USB port.
Returns the bus address of the instrument; This is the address used by
GPIB, if fitted, or may be used as a general identifier over the other
interfaces.
The PSU has no self test capability and the response is always 0 <rmt>
The PSU has no trigger capability. The command is accepted but performs
no action.