PROBLEM
GelPad contact symbol appears
on the display (
on the display (
)
)
GelPads don't adhere to body
Unpleasant feeling beneath the
GelPads
Error message E001, E002, E003
appears on the display
GELPAD CONTACT INDICATOR (
GELPAD CONTACT INDICATOR (
Your
5 unit has a
SLENDERTONE REVIVE S
between the GelPads and your skin. When poor contact is detected:
• The intensity bar for the side(s) affected will flash.
• The GelPad contact symbol will begin flashing on the display.
• Your unit will beep and the timer will pause.
• The intensity level will drop to zero and the up intensity keys will be disabled.
If this happens, you should check that the GelPads and leads are properly connected and that all GelPads are pressed fully against your skin.
If this doesn't work, check that the leads are properly connected to the unit. When proper contact is restored, stimulation builds over a 2 sec-
ond period to the previously set intensity level.
www.slendertone.com
English
POSSIBLE CAUSES
Lead(s) not fully inserted in sockets
GelPads not connected to leads
GelPads are worn
GelPads are worn
Check the skin for lotions, creams,
cosmetics etc.
Poor GelPad contact
Battery is low
Unit is faulty
)
)
GelPad
contact feature which continuously checks the connection between the leads and GelPads and
SOLUTIONS
Re-insert leads correctly
Reconnect the GelPads
Visit the website or contact your local
retailer for new GelPads
Visit the website or contact your local
retailer for new GelPads
Clean the skin thoroughly
Ensure GelPads are flat against your skin
Replace the battery as per page 9
Return the unit for repair or replacement