Problem
Mask is too noisy

Technical Specifications

Note: The mask system does not contain latex, PVC or DEHP materials.
Pressure–Flow Curve
70.0
60.0
50.0
40.0
30.0
20.0
10.0
0.0
2
4
6
8
Mask Pressure (cm H
Note: As a result of manufacturing variations, the vent flow may vary from that
indicated in the graph.
Dead Space Information
Dead space is the empty volume of the mask up to the elbow.
The maximum dead space of the Mirage Liberty for all mask size combinations
is 151 mL.
Therapy Pressure
4 to 20 cm H
O
2
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.5 cm H
at 100 L/min: 1.6 cm H
Possible reason
Vents are blocked or partially
blocked.
Valve is misassembled.
Cushion is not properly inserted. Check insertion of cushion.
Mask is assembled incorrectly.
Valve is damaged, torn or
distorted.
Pillows are not inserted correctly
into cushion.
Mask is not sealing.
10
12
14
16
18
20
22
24
O)
2
O
2
O
2
Possible solution
Clean vents according to
instructions in "Cleaning the
Mask in the Home" .
Reassemble valve.
Reassemble mask.
Replace valve.
Check insertion of pillows.
Check and adjust mask as
necessary.
Pressure
(cm H
O)
2
4
8
12
16
20
26
28
30
8
Flow
(± 6.0 L/min)
22
32
41
48
54