8  Maintaining your STATIM
8.6  Changing the Bacteria Retentive Air Filters
The filters should be replaced every six months or after 500 cycles to maintain
an adequate supply of clean air during the air drying cycle.
To change the bacteria retentive air filter on the STATIM 2000 G4 and 5000 G4,
follow these steps:
1. Power the STATIM OFF.
2. Disconnect tube A
retentive filter
the filter bracket
from the bracket, note the orientation of the
arrow mark on the filter.
3. When the filter is free of the bracket, carefully
disconnect tube B      from the filter.
4. Before installing the replacement
bacteria retentive filter       (SciCan order
no. 01-102119S) check that the arrow mark
on the filter matches the direction
of the arrow on the bracket. Push the left
hand filter fitting into tube B
5. Gently press the replacement filter into
the filter bracket
the filter should be facing out and pointing to
the left.
6. Re-connect tube A      to the right hand filter fitting.