Measurement parameters
'
Mean Lambda (λ
) Represents the center of mass for the trace points.
Peak lambda
Wavelength of the FP source peak spectral component.
Mode Spacing
Is the average wavelength spacing between the individual spectral
components of the FP source.
FWHM
(Full Width at Half Maximum) describes the spectral width for the
FP source half-power (-3dB) points, assuming a continuous,
Gaussian power distribution based on defined λ
Power
Result
represented.
Peak Amplitude
Power level for the FP source peak spectral component.
Sigma (σ)
Equivalent value of FP source spectral width for a Gaussian
distribution based on the previously defined λ
File
Data file menu access.
Options
Save Values
Manage Files
Select
Delete
Load
Close Menu
Exit
Page 30
of
energy
integration
Saves in a file the registered measurement parameters
corresponding to the visualised trace.
Shows the data file list saved in memory.
Selects
a
data
parameters previously described.
It allows deleting a data file.
Load a data file.
Quits data file menu.
Quits measurement function.
USER'S MANUAL. PROLITE-60
and σ.
'
over
the
wavelength
'
.
file
containing
the
range
measurement
09/2007