8.3.4 Saving and selecting the program

By touching the program name on the main side opens a
program selection menu.
Four factory programs (e.g. Gastro F3,2) are set, whose
parameters can be changed. The factory settings are
retained when the device is switched off and on again.
For example, four client-programs are entered, whose
parameters can be set. After shut down, the last set values
are saved.
8.4

Operating errors

Following Error messages can occur:
Error
11: "EEROR in CPU
ROM"
12: "EEROR in CPU
RAM"
13: „ARGON level
LOW"
14: „ARGON no
FLOW"
38
Localization
Explanation
CPU_Failure
Wrong check sum of the ROM
– Memory of the control
processor
Wrong check sum of the RAM
– Memory of the control
processor
ARGON_Block
The pressure in the cylinders
is below the required
minimum or there is an error
in the pressure measurement
system
The flow volume set at the
display cannot be reached
HBS Argon
Cause
Controller
Controller
Argon module
Argon module