Oszillo-saw GOLD II accu autopsy
LED indicator
1 – 4 green LED
Continuous red light
Red flashing light
The electronics automatically switch off the motor prior to the battery being deep discharged.
Operate and charge the battery only within the battery operating-temperature range of 0 °C – 45
°C (32 °F – 113 °F). At the beginning of the charging procedure, the battery temperature must be
within the battery operating- temperature range.

5 Mounting / exchanging the saw blade

!

5.1 Releasing the retaining screw / removing the old saw blade

1.
Place the cast cutting saw on a stable underlay (table).
2.
Apply one fork wrench at the drive shaft behind the saw blade, and the other one at the front at the
retaining screw (see
3.
Hold the rear wrench firmly and use it to fix the cast cutting saw.
4.
Release the retaining screw using the front wrench.
5.
Remove the retaining screw and the fixing ring. Ensure that you do not lose any components.
6.
If applicable, remove the old saw blade.
Meaning
Percentage of charge condition
Battery is almost empty
Battery is not ready for
operation
Danger of injury when unintentionally turned on. Remove
accumulator before changing. Should it be necessary, use stable
gloves so that you Can not be injured by the saw blades. Only
connect the accumulator when the saw is switched off.
).
fig. 3
Activity
Operation
Charge battery
Bring the battery into the battery
operatingtemperature range, then
charge
EN
21