1. Applications

The compact SE 610 conductivity sensor has coaxially arranged electrodes
made of stainless steel. A temperature detector is integrated in the sensor.
Electrical connection is made via a 5 m long fixed cable. Its low costs, the
fast reacting temperature detector, the broad measurement range, and its
robust design make the sensor particularly suitable for the following appli-
cations: Drinking water, industrial wate, surface water, ion exchangers and
reverse osmosis plants, rinse water, seawater desalination plants.

2. Dimension Drawing

3. Calibration / Adjustment
Remove sensor, clean, rinse, and dry it, then immerse it in a calibration solu-
tion. Suitable calibration solution ZU 0350 (15 µS/cm), ZU 0702 (147 µS/cm).
Perform a product calibration if the sensor can or shall not be removed.
2
G3/4A
G1/2A
Ø 14
Connecting cable
Anschlussbild Kabel
Connection
SW 27 PEI
Electrodes 1.4571
braun
LF
weiss
grün
Temp.
gelb