Call up menu:
 Press the Menu button to access the menu and
the first menu item "PRoGRM" will be dis-
played.
 Use navigation buttons   to select menu
item "CAL".
 Acknowledge using ENTER key, the current
setting is displayed. If necessary, use naviga-
tion buttons   to select "TEMP".
 Press ENTER to confirm and the first tempera-
ture point to be tested will be displayed.
 Press START and the first heating-up period
will start.
• Do not leave the heated cover open for more than 1 minute during
the heating-up period, otherwise the error message "ERR.100" will
appear. In this case press ESC and restart menu item "TEMP".
• Attention: some parts e.g. ventilation grilles and inspection windows
may become very hot during operation.
• The adjustment must be made within 15 minutes when the tempera-
ture display flashes, otherwise the temperature adjustment will be
aborted (AboRT will be displayed).
 Temperature calibration for the first point takes
15 min. Compare the displayed temperature
value on the thermometer with that of the mois-
ture analyser whilst the temperature display is
flashing. If the two values do not match, correct
them using the navigation buttons   and
confirm with the ENTER button. Numeric input
see operating instructions KERN DBS, chap.
7.1.1.