Call up menu:
Press the Menu button to access the menu and
the first menu item "PRoGRM" will be dis-
played.
Use navigation buttons to select menu
item "CAL".
Acknowledge using ENTER key, the current
setting is displayed. If necessary, use naviga-
tion buttons to select "TEMP".
Press ENTER to confirm and the first tempera-
ture point to be tested will be displayed.
Press START and the first heating-up period
will start.
• Do not leave the heated cover open for more than 1 minute during
the heating-up period, otherwise the error message "ERR.100" will
appear. In this case press ESC and restart menu item "TEMP".
• Attention: some parts e.g. ventilation grilles and inspection windows
may become very hot during operation.
• The adjustment must be made within 15 minutes when the tempera-
ture display flashes, otherwise the temperature adjustment will be
aborted (AboRT will be displayed).
Temperature calibration for the first point takes
15 min. Compare the displayed temperature
value on the thermometer with that of the mois-
ture analyser whilst the temperature display is
flashing. If the two values do not match, correct
them using the navigation buttons and
confirm with the ENTER button. Numeric input
see operating instructions KERN DBS, chap.
7.1.1.