Repairs / Decommissioning / Disposal

10 Repairs

The sensor can be repaired only at the factory. Send it to us in its original
packaging. Prepare it for the repairs (as described in chapter 11

"Decommissioning").

11 Decommissioning
Decommission the sensor: follow all of the safety instructions in chapter 5
"Assembly".

12 Disposal

Electrolyte
The electrolyte can be poured into a drain.
Sensor
30
Disconnect the sensor from the electricity supply (compare chapter 6
"Installation")
Depressurise the in-line probe housing
Loosen the terminal screw
Pull the sensor out of the in-line probe housing slowly
Unscrew the membrane cap over a sink and empty it
Throw away the membrane cap
Thoroughly remove all traces of gel from the electrodes using clean, warm
water and allow to dry free of dust
Screw on a new membrane cap loosely to protect the electrodes
Put on the membrane's protective cap to protect the membrane cap
IMPORTANT
• Electronic rubbish is classed as hazardous waste.
• Please comply with the current regulations in your
country.
®