Télécharger Imprimer la page

OHAUS PAJ Série Manuel D'instruction page 7

Masquer les pouces Voir aussi pour PAJ Série:

Publicité

Les langues disponibles

Les langues disponibles

PAJ-N Series
^^^^^^^^^^^^^^^^^^^^^
User ID: ............................
Bal ID: 123456789
Proj ID: ...........................
Time: ......:.......:.......
Date:
...../ ..... / .....
120.01 g
^^^^^^^^^^^^^^^^^^^^^
4.7 RS232 Menu [.rs232.]
Baud [BAUD] - The RS232 baud rate can be set to 600, 1200, 2400, 4800, 9600 and 19200.
Parity [Parity] - Parity can be sent to 7 bits-even parity [7 eveN], 7bits-odd parity [7 odd], 7bits-no parity
[7 No] or 8bit-no parity [8 No].
Handshake [Handsh] - Handshake can be set to off [set Off], X on–X off [on-off], or hardware [HardwWr].
4.8 Lock Menu [.LOC.]
When a Lock Menu item is set on the indicated menu cannot be changed.
[Loc Cal] - Calibration menu lock
[Loc.Unit] - Unit menu lock
5.0 LEGAL FOR TRADE
When the balance is used in trade or legally controlled applications it must be set up, verified and sealed in
accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that
all pertainent legal requirements are met.
5.1 Settings
Before the verification and sealing perform the following steps:
1. Verify the menu settings meet the local weights and measures regulations.
2. Perform a calibration as explained in Section 2.5.
3. Set the position of the security switch as shown in Section 5.3.
Note: When the security switch is set on the following menu settings cannot be changed:
CAL, SETUP, MODE, UNIT and LOC.
5.2 Verification
The local weights and measures official or authorized service agent must perform the verification procedure.
5.3 Sealing Access to the Balance Settings
The local weights and measures official or authorized service agent must apply a security seal to prevent
tampering with the settings. Refer to the illustrations below for the sealing methods.
Un-Locked
GLP (ON)
Balance ID (ON)
GLP (ON)
GLP (ON)
GLP (ON)
Result
Multi Line Format with 4 line feeds
Line Feed-2
Line Feed-3
Line Feed-4
[Loc SEt] - Setup menu lock
[Loc Prt] - Print menu lock
Locked with Paper Seal
EN-6
[Loc.mMod] - Mode menu lock
[Loc 232] - RS232 menu lock
Locked with Wire Seal

Publicité

loading