Manual english
6.
MAINTENANCE
6.1
Sensor cleaning
Cleaning of the measuring head is recommended before emerging in other oil and prior to calibration.
Do not touch or rub the sensing element during the cleaning process!
Cleaning of sensor from oil residue:
1) Emerge the measuring head in N-HEPTAN and swirl for approx. 30 seconds
2) Remove excess liquid and allow to air dry for approx. 30 minutes
!
Warning:
It is strongly advised NOT to try cleaning the sensor by mechanical means, such as rubbing the sensor with cotton material!
6.2
Self-diagnostic and error messages
Self-diagnostic via LED on the circuit board:
Self-diagnostic via display (option):
Power LED green
LED D1 blue
page 25
Power LED (green):
- fl ashing
=> Supply voltage applied / Microprocessor is active
LED D1 (blue):
- constantly lit => sensor element damaged
- fl ashing
=> sensor element bedewed (condensation)
- Error 1
=> Humidity sensor element damaged
- Error 2
=> Humidity sensor element bedewed (condensation)
- Error 3
=> Temperature sensor element damaged
- Error 4
=> Short circuit of temperature sensor
EE381