Manual english
4.
ELECTRICAL CONNECTIONS
Male connector:
5.
MAINTENANCE
5.1
Sensor cleaning
Cleaning of the measuring head is recommended before emerging in other oil and prior to calibration.
Do not touch or rub the sensing element during the cleaning process!
Cleaning of sensor from oil residue:
1) Emerge the measuring head in N-HEPTAN and swirl for approx. 30 seconds
2) Remove excess liquid and allow to air dry for approx. 30 minutes
!
Warning:
It is strongly advised NOT to try cleaning the sensor by mechanical means, such as rubbing the sensor with cotton material!
analogue output
4
3
1...V+
2...output 1
1
2
3...GND
4...output 2
page 15
EE385