Digital Water BathS
SWB6D, SWB15D and SWB24D
Instructions for use
introduction
Thank you for purchasing this piece of Stuart
equipment. To get the best performance from the
equipment and for your own safety, please read
these instructions carefully before use. Before
discarding the packaging check that all parts are
present and correct.
This equipment is designed to operate under the
following conditions:
 For indoor use only
 Use in a well ventilated area
 Ambient temperature range +5ºC to +40ºC
 Altitude to 2000m
 Relative humidity not exceeding 80%
 Mains supply fluctuation not exceeding 10%
 Over-voltage category II IEC60364-4-443
 Pollution degree 2
If the equipment is not used in the manner
described in this manual and with accessories other
than those recommended by Stuart, the protection
provided may be impaired.

electrical installation

thiS eQUiPMeNt MUSt Be eartheD
Before connection please ensure that the line
supply corresponds to that shown on the rating
plate.
Model
Supply requirements
SWB6D
230V, 50Hz
SWB15D
230V, 50Hz
SWB24D
230V, 50Hz
All models are supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One has a U.K. 3 pin plug and the other has a 2 pin
"Shuko" plug for connection to the mains. Choose
the lead appropriate for your electrical installation
and discard the other. Should neither lead be
suitable, take the lead with the U.K. plug and
replace the plug with a suitable alternative. This
involves cutting off the moulded plug, preparing the
cable and connecting to the re-wireable plug in
accordance with its instructions.
thiS OPeratiON ShOUlD ONlY Be
UNDertaKeN BY a QUaliFieD eleCtriCiaN
NOte: Refer to the equipment's rating plate to
ensure that the plug and fusing are suitable for the
voltage and wattage stated
Fuses
230V: 2 x 10A fuses. For all other voltages always
replace with an equivalent fuse type and rating as
those being removed, alternatively see rating plate
for fuse type and rating.
The wires in the mains cable are coloured as follows:
BROWN
BLUE
GREEN/YELLOW
The instruments are fitted with an IEC socket at the
rear of the instrument for connection of the mains
lead. The appropriate mains lead should be
connected BeFOre connection to the mains supply.
When wishing to isolate the bath from the mains
supply, disconnect the cable from the IEC socket.
Should the mains lead need replacement, a cable of
1mm
IEC320 plug should be used.
iF iN DOUBt CONSUlt a QUaliFieD
eleCtriCiaN
Power
350W
750W
1400W
CaUtiON – Fuse fitted in both live and
neutral lines.
-
LIVE
-
NEUTRAL
-
EARTH
2
of harmonised code H05W-F connected to an
1