ANOMALIES AND SOLUTIONS

ANOMALIE
Loss of water
The device does not provide water
The machine stops and does not
have time to fill
The autoclave indicates that the tank
is full of water used
KIARA-TEC /manuale di istruzioni-IT/ Rev.02/08.07.14
POSSIBLE CAUSE
2.
Fittings not properly closed
3.
Solenoid valve stuck
4.
Water pressure too low
2.
It is' set little time for the
filling
2.
Possible obstruction of the
discharge from the autoclave
and siphon
POSSIBLE SOLUTIONS
2.
Proceed to fasten all
connections
3.
Try to unlock the valve
4.
Increase the pressure
2.
Add a few minutes to the
duration of opening of the
valve
2.
Proceed to clear the drain.
Clean the siphon.