INSTRUCTIONS FOR INSTALLATION
A. Carefully remove the packaging and save all packing material.
B. Place the Kiara-Tec preferably under the sink and near the sterilizer.
C. Make sure, with a pressure gauge, that the water pressure is not less than 2 bar, constant throughout the year.
D. Connect the TAP load (5282) to a drinking water intake, using the kit provided.
E. Connect the DRAIN (5485) after the trap of the sink.
F. Connect the POWER (5262) to the sterilizer
G. Turn on the sterilizer and set the automatic loading (see page 20. The manual use of the sterilizer)
H. Connect the ATTACK (5337) to the black socket of the sterilizer
I. Place the attack supply (5337) in a container on the floor
J. Open the valve (5282) and check for leaks
K. Try to choose a sterilization cycle
L. If everything is successful, from the attack (5337) starts out a bit 'of water. The conductivity monitor the quality
to be below 15 microsiemens. Take out about 15 liters of water, this procedure may last for about 30 minutes.
M. Connect the attack (5337) to the sterilizer (white).
N. Check the value of water directly on the display while running the autoclave loading
Besides minimal equipment, the technician must have the conductivity for the immediate control of
water quality!
Warning! The drain hose must be long enough and should have a downward path.
Warning! The drain siphon must be free flowing!
Warning!
Look out! This osmosis system to produce 1 liter of water, will eliminate about 5 liters.
KIARA-TEC /manuale di istruzioni-IT/ Rev.02/08.07.14