boso medicus family Mode D'emploi page 69

Masquer les pouces Voir aussi pour medicus family:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Calibration checks - Testing instructions
(for trained specialist staff only)
Testing
1.) Remove batteries.
2.) Hold start/stop button
down and insert batteries.
3.) Release start/stop button
The monitor is now ready for testing;
the current pressure is displayed in the
SYS, DIA and PULS fields. The
start/stop button
green.
4.) Perform test for deviation of pressure
display and pressure circuit integrity
5.) Switch off monitor by pressing the
pressed
.
C) Safety seal
is illuminated
(observe setting time for cuff – at least
30 seconds) in the usual manner.
start/stop button
establish all original connections.
As a safeguard, the upper and lower
parts of the housing can be joined
with a safety seal. Alternatively, a hole
drilled in the bottom of the monitor for
the connecting screw for the housing
can be sealed for safety.
again and re-
69

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ca01Ca02Ca04

Table des Matières