9. Specifications
Limited error with selected temperatures for thermocouples types K and N
Temperature in
(ITS-90)
0
[32]
EN
100
[212]
200
[392]
300
[572]
400
[752]
500
[932]
600
[1,112]
700
[1,292]
800
[1,472]
900
[1,652]
1,000 [1,832]
1,100 [2,012]
1,200 [2,192]
Other thermocouple types on request
For further specifications see WIKA data sheet TE 65.60 or Technical information IN 00.23
and the order documentation.
20
Tolerance value IEC 60584-1 in
°C [°F]
Class 1
±1.5 [±34.7]
±1.5 [±34.7]
±1.5 [±34.7]
±1.5 [±34.7]
±1.6 [±34.88]
±2.0 [±35.6]
±2.4 [±36.32]
±2.8 [±37.04]
±3.2 [±37.76]
±3.6 [±38.48]
±4.0 [±39.2]
-
-
WIKA operating instructions model TC59-T (TEFRACTO-PAD
°C [°F]
Class 2
±2.50 [±36.5]
±2.50 [±36.5]
±2.50 [±36.5]
±2.50 [±36.5]
±3.00 [±37.4]
±3.75 [±38.75]
±4.50 [±40.1]
±5.25 [±41.45]
±6.00 [±42.8]
±6.75 [±44.15]
±7.50 [±45.5]
±8.25 [±46.85]
±9.00 [±48.2]
)
®