Table 4
Recommended separation distances between portable and mobile RF communications equipment and the BM 35
The BM 35 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
BM 35 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the BM 35 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from structures,
objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
Not applicable
–
–
–
–
–
79
(m)
80 MHz to 800 MHz
d = 1,2
0,12
0,38
1,2
3,8
12
800 MHz to 2.5 GHz
d = 2,3
0,23
0,73
2,3
7,3
23