4.5
Sample "Water Monitoring in µS Range"
Settings:
Unit configuration:
Controller configuration:
Unit parameter assignment:
Controller parameter
assignment
Limit contact:
NOTE
ATTENTION
417101802_Anhang02_LF_konduktiv.doc
Conductivity Measuring Module (contact measurement)
Current output
Temperature unit
Temperature compensation
Temperature coefficient
Signal unit:
Direction of action
External control enabling
Measuring range:
Calibration must be carried out
Values for conductivity min/max
can be adjusted optionally
Adjust setpoint value W to e.g.
B.95 µS
Switch difference XSD
ON delay
OFF delay
Switching point distance LW
Switch difference X2SD
Limit contact L-:
Limit contact L+:
Switch difference X2SD
The 1/2 relay output terminal is designed as an alarm output for the limit contact.
The 3/4 and 5/6 output terminals are always assigned to measuring module 1. Relay
output terminals for measuring modules 2 and 3 are located on relay board 255119.
The current output terminals are also assigned to the modules. These output
terminals are galvanically isolated in the case of base board
(material-no. 35514004).
When connecting the current output terminals, pay attention to the polarity and the
maximum load (600 Ω).
- 46 -
(0/4..20mA) optional
Celsius
auto
corresponds to product
data sheet or self-
determined
positive
yes
0..200 μS
2 %
0 sec.
0 sec.
N/A
N/A
e.g. 50 μS
e.g. 150 μS
2 %
Rev.: 05-07/10