The selected measuring range must first of all be checked and possibly changed.
Value range: 0..2 µS / 0..20 µS / 0..200 µS
NOTE
CAUTION
The continue function cancels the calibration and opens the calibration module selection
menu
The calibration start function starts the conductive C calibration.
The continue function cancels the calibration and opens the calibration module selection
menu.
Probe in air
The measuring probe must now be placed outside of a conductive environment. Once this
has been ensured, the calibration can be continued through selecting ENTER.
Measurement 1
Measurement 1 takes place automatically and continues until the measurement range is
stable.
Introduce simulation
resistance
417101802_Anhang02_LF_konduktiv.doc
Conductivity Measuring Module (contact measurement)
It must always be ensured that the correct measuring range has been selected, as
failure to do so can lead to erroneous measurements (overflow if the measuring
range is too small) or inaccuracies (measuring range too large).
Increasing error is to be expected in the case of measurements below 0.5 µS /cm.