3
Connection
3.1
Electrical connection
The measuring probes are equipped with 10 m of cable at the factory. A special shielded
cable (Material-No.418437041) and a special terminal box (Material-No. 288101) must
be implemented if larger distances are to be bridged.
CAUTION
Electrode cables must not be placed into one cable duct together with mains leads.
Fig. 3.1
Terminal connection diagram
Contact controller enabling closed ⇒ controller active
417101802_Anhang02_LF_konduktiv.doc
Conductivity Measuring Module (contact measurement)
The maximum length of the cable must not exceed 100 m
- 30 -