Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the SBM 03
The SBM 03 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of the SBM 03 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the SBM 03 as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum output
power of transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be deter-
mined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
Not applicable
58
(m)
80 MHz to 800 MHz
d = 1,2
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2,3
0,23
0,73
2,3
7,3
23