en - Areas of application
Loading mesh
trays with injector
manifolds
34
Place the mesh trays in mesh tray carrier A 311. The water connec-
tions must point towards the injector manifold.
Load from the back to the front, and unload from front to back.
The mesh trays must not be placed loosely in the mobile unit or
stacked on top of each other.
Connect the mesh trays to the A 207 mobile unit with the silicone
hoses.
Secure the ends of any unused hoses in the perforated plate.
Otherwise unused silicone hoses will be spun around by the
force of the water. This could cause damage to the load or to the
machine.
Unused hoses must not be sealed with stoppers.