Loading lumened
instruments
Loading mesh
trays without in-
jector manifolds
Arrange the instruments in the mesh trays according to their manu-
facturer's instructions.
When loading the instruments the following must be observed:
1. When processing hollow instruments with several connectors, such
as rinsing and suction hand pieces, each connector needs to be
secured to a hose connector in the mesh tray.
2. Lightweight and delicate instruments must be secured in silicone or
plastic holders. Space the holders in the mesh tray according to the
size of the instruments.
Arrange the mesh trays as illustrated in the lower levels of the A 207
mobile unit.
The mesh trays must not be stacked on top of each other.
en - Areas of application
33