SPECIFICATIONS

Catalog Number
Delivery Rate
Maximum Recommended Flow (@
nominal outlet pressures of zero and
7 kPa)**
Outlet Pressure
Auxiliary Oxygen Port **
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 70˚F {21˚C} only)
0-1500 M (0-5000 ft)
Operating Environment Range*
41˚F (5˚C) to 95˚F (35˚C), humidity
range of 15% to 93% non-
condensing
Power Consumption
Weight
Safe Working Load
Sound Pressure Level at 3 and 10
LPM
Sound Power Level at 3 and 10 LPM
Dimensions
Maximum Limited Pressure
Operating System
Low Oxygen Indicator
Storage Conditions
Equipment Class and Type
Approval Body and Safety Standard
CE mark
EMC Compliance To
*
NOTE– The OSD performance at 41˚F (5˚C) to 95˚F (35˚C), 93% R.H. through voltage range on the 1025 verified at 670m.
**The maximum recommended flow is 6 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.
Specifications subject to change without notice.
SE-1025-1
DEVILBISS 10-LITER SERIES
1025DS
2 to 10 LPM
10 LPM
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
120 V, 60 Hz, 6 .1 Amp
102-132 V~, 60 Hz
2-10 LPM=87%-96%
Across the voltage range:
No degradation of performance
No degradation in performance across
the operating voltage range .
120 vac, 60HZ: 639 watts average
42 lbs. (19 Kilograms)
53 lbs. (24 Kilograms)
<59 dBA
<69 dBA
24.5"H x 13.5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IP21
TUV
ANSI/AAMI ES60601-1:2005+A2
(R2012) +A1
IEC 60601-1-6:2010
IEC 60601-1-11:2015
ISO 80601-2-69:2014
CAN/CSA-C22.2 No. 60601-1:14
CAN/CSA-C22 .2 No . 60601-1-6:11
CAN/CSA-C22 .2 No . 60601-1-11:15
CAN/CSA-C22 .2 No . 80601-2-69:16
No
EN60601-1-2
1025KS
2 to 10 LPM
10 LPM
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
230 V~, 50 Hz, 3 .2 Amp
195-253 V~, 50 Hz
2-10 LPM=93% +3%/-6%
Across the voltage range:
No degradation of performance
No degradation in performance across
the operating voltage range .
230 vac, 50Hz: 664 watts average
42 lbs. (19 Kilograms)
53 lbs. (24 Kilograms)
<57 dBA
<67 dBA
24.5"H x 13.5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part
IP21
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
Yes
EN60601-1-2
1025UK
2 to 10 LPM
10 LPM
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
240 V~, 50 Hz, 3.2 Amp
204-264 V~, 50 Hz
2-10 LPM=93% +3%/-6%
Across the voltage range:
No degradation of performance
No degradation in performance across
the operating voltage range .
240 vac, 50Hz: 670 watts average
42 lbs. (19 Kilograms)
53 lbs. (24 Kilograms)
<57 dBA
<67 dBA
24.5"H x 13.5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part
IP21