Technical Specification
Rotational speed
Start-up time
Pwr supply
(Line type):
Input voltage:
Input freq.:
Max input pwr:
Stand-by pwr:
Max out power:
Pwr supply
(24 Vdc type):
Input voltage:
Max input pwr:
Stand-by pwr:
Max out power:
Protection fuse
Protection fuse
Compliance with:
Power cable
Serial communica-tion ( kit)
Operating temperature
Storage temperature
Pump Output
Controller TV70 Navigator User Manual / 87-900-956-01 (G)
75 Krpm (default and Max)
setable from 75 down to 15 Krpm
< 2 minutes
100 - 240 Vac
50 - 60 Hz
210 VA
10 W
75 W in normal air or water cooling
100W in Ramp air or water cooling
24 Vdc (20 - 28 Vdc)
140 W
10 W
75 W in normal air or water cooling
100W in Ramp air or water cooling
4 A (Line type):
8 A (24 Vdc type):
EN-CENELEC 55011
IEC 1000-4-2 (ex 801-2)
IEC 1000-4-3 (ex 801-3)
IEC 1000-4-4 (ex 801-4)
EN 61010-1 (IEC 1010-1)
With European or NEMA plug, 3 meters long (optional)
RS232 cable with a 9-pin D type male connector and a 9-pin D type
female connector, and software (optional)
0° C to +40° C
-20° C to +70° C
54 Vac, three phase,
1250 Hz (max)
Technical Information
Technical Specification
18
151/182