Problem
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD shows "Er 8"
LCD shows "Er A"
No response
when you press
a button or load
batteries.
1.
Do not drop this monitor nor subject it to strong impacts.
2.
Avoid high temperatures and direct sunlight. Do not immerse
the monitor in water as this will result in damage to the monitor.
3. If this monitor was stored at a temperature near the freezing
point, allow it to come to room temperature before use.
4.
Do not attempt to disassemble this monitor.
5. If you do not use the monitor for a long time, please remove the
batteries.
162
PC-BMG3018_IM
Possible cause
Cuff pressure above
300 mm Hg
More than 3 minutes
with cuff pressure above
15 mmHg
EEPROM accessing
error
Appliance parameter
checking error
Pressure sensor param-
eter error
Incorrect operation or
strong electromagnetic
interference.
Maintenance
Solution
Measure
again after
five minutes. If
the monitor is
still abnormal,
please contact
the local
distributor or the
factory.
Take out
batteries for five
minutes, and
then reinstall all
batteries.
04.07.18